Immunoturbidimetry Reagents

Immunoturbidimetry is an important technique in the diagnostic field of clinical chemistry.

It is used to determine serum proteins that are not detectable with classical clinical chemistry methods and antigen-antibody reaction.

CCR provides immunoturbidimetric tests in various kits for multi-purpose or automated use on common clinical chemistry analyzers.

Download PDF

nimble_asset_PDF

IMMUNOTURBIDIMETRY

(Application procedures on clinical chemistry analyzers are available upon request.)

α1-ACID GLYCOPROTEIN

Quantitative determination of a1-Acid Glycoprotein (AGP) in human serum by turbidimetric immunoassay. Measurement of antigen-antibody reaction by the end-point method.

Liquid reagents, ready for use.

Measuring Range : 0 – 300 mg/dL  Detection Limit : 4 mg/dL.

REF IMT279-K080  R1 3x25ml   R2 1x25ml

ANTITHROMBIN III

Quantitative determination of  Antithrombin III (AT3) in human plasma by turbidimetric immunoassay. Measurement of antigen-antibody reaction by the end-point method.

Liquid reagents, ready for use. Measuring Range : 0 – 65 mg/dL. Detection Limit : 8 mg/dL.

REF IMT276-K080  R1 3x25ml   R2 1x25ml

ANTI-STREPTOLYSIN O (ASO)

Reagents for measurement of ASO concentration. Serum anti-streptolysin O (ASO) causes agglutination of latex particles coated with streptolysin O. The agglutination of the latex particles is proportional to the streptolysin O concentration and can be measured by turbidimetry. Linearity limit: 800 IU/mL ASO. Detection limit: 3 IU/mL ASO.

REF IMT481-K050  R1 2x20ml   R2 1x10ml + CAL 1x1ml

COMPLEMENT C3

Quantitative determination of Complement C3 (C3C) in human serum by turbidimetric immunoassay.

Measurement of antigen-antibody reaction by the end-point method.

Liquid reagents, ready for use.

Measuring Range : 0 – 400 mg/dL  Detection Limit : 20 mg/dL.

REF IMT277-K065  R1 3x20ml   R2 1x5ml

COMPLEMENT C4

Quantitative determination of Complement C4 (C4C) in human serum by turbidimetric immunoassay.

Measurement of antigen-antibody reaction by the end-point method.

Liquid reagents, ready for use.

Measuring Range : 0 – 80 mg/dL  Detection Limit : 2 mg/dL.

REF IMT278-K065  R1 3x20ml   R2 1x5ml

C-REACTIVE PROTEIN (CRP)

Reagents for measurement of CRP concentration. Serum C-reactive protein (CRP) causes agglutination of the latex particles coated with antihuman C-reactive protein. The agglutination of the latex particles is proportional to the CRP concentration and can be measured by turbidimetry. Linearity limit: 150 mg/L CRP. Detection limit: 1.0 mg/L CRP.

REF IMT483-K050  R1 2x20ml   R2 1x10ml + CAL 1x1ml

DIRECT HbA1c

Reagents for measurement of Hemoglobin A1c concentration.

After preparing the hemolysate, the Hemoglobin A1C (HbA1C) concentration is quantified by a latex turbidimetric assay. Reagents are provided ready to use.

Auxiliary Reagents supplied separately:  IMT137-CAL002  CALIBRATOR SET – IMT128-CTR001 CONTROL SET

Measuring Range : 6 – 140 mmol/mol. Detection limit: 6 mmol/mol.

REF IMT458-K036  R1 1x30ml   R2 1x6ml R3 2×60 ml lysing solution.

FERRITIN

Reagents for measurement of ferritin concentration. Working Reagent: pour the contents of a Reagent B vial, into a Reagent A bottle. Mix thoroughly. Stable for 30 days at 2-8ºC. Measurement interval: 4-500 µg/L. Detection limit: 4 µg/L ferritin. The ferritin standard is included in the package.

REF IMT584-K045  R1 2x15ml   R2 1x15ml + CAL 1x3ml

FIBRINOGEN

Quantitative determination of  Fibrinogen (FIB) in human plasma by turbidimetric immunoassay. Measurement of antigen-antibody reaction by the end-point method. Liquid reagents, ready for use.

Auxiliary Reagents supplied separately:  IMT780-STD0000.5 Fibrinogen standard.

Measuring Range : 0 – 523 mg/dL  Detection Limit : 4.5 mg/dL.

REF IMT289-K080  R1 3x25ml   R2 1x15ml

IgA

Quantitative determination of IgA in human serum by turbidimetric immunoassay. Measurement of antigen-antibody reaction by the end-point method. Liquid reagents, ready for use.

Measuring Range: 0 – 600 mg/dL. Detection Limit: 1 mg/dL

Auxiliary Reagents supplied separately

– saline solution 0.9g/dl

– set of calibrators to prepare the 5-level curve: IMT796-STD005 Protein standard set, or alternatively- high concentration calibrator, to generate a reference curve: IMT794-STD001 Protein standard high– CONTROLS: IMT805-CTR001 Himmunology control lowIMT806-CTR001 Immunology control high.

REF IMT282-K045  R1 2x20ml   R2 1x5ml

IgG

Quantitative determination of IgA in human serum by turbidimetric immunoassay. Measurement of antigen-antibody reaction by the end-point method. Liquid reagents, ready for use.

Measuring Range: 0 – 2700 mg/dL. Detection Limit: 90 mg/dL

Auxiliary Reagents supplied separately

– saline solution 0.9g/dl

– set of calibrators to prepare the 5-level curve: IMT796-STD005 Protein standard set, or alternatively- high concentration calibrator, to generate a reference curve: IMT794-STD001 Protein standard high– CONTROLS: IMT805-CTR001 Himmunology control low

IMT806-CTR001 Immunology control high.

REF IMT281-K045  R1 2x25ml   R2 1x5ml

IgM

Quantitative determination of IgA in human serum by turbidimetric immunoassay. Measurement of antigen-antibody reaction by the end-point method. Liquid reagents, ready for use.

Measuring Range: 0 – 500 mg/dL. Detection Limit: 6 mg/dL

Auxiliary Reagents supplied separately

– saline solution 0.9g/dl

– set of calibrators to prepare the 5-level curve: IMT796-STD005 Protein standard set, or alternatively- high concentration calibrator, to generate a reference curve: IMT794-STD001 Protein standard high– CONTROLS: IMT805-CTR001 Himmunology control low

IMT806-CTR001 Immunology control high.

REF IMT283-K045  R1 2x20ml   R2 1x5ml

KAPPA LIGHT CHAIN

Quantitative determination of  Kappa Light Chain (KAP) in human serum by turbidimetric immunoassay. Measurement of antigen-antibody reaction by the end-point method. Liquid reagents, ready for use. Measuring Range: 0 – 800 mg/dL. Detection Limit: 40 mg/dL

Auxiliary Reagents supplied separately

– saline solution 0.9g/dl

– set of calibrators to prepare the 5-level curve: IMT796-STD005 Protein standard set, or alternatively- high concentration calibrator, to generate a reference curve: IMT794-STD001 Protein standard high– CONTROLS: IMT805-CTR001 Himmunology control low

IMT806-CTR001 Immunology control high.

REF IMT286-K080  R1 2x25ml   R2 1x5ml

LAMBDA LIGHT CHAIN

Quantitative determination of  Lambda Light Chain (KAP) in human serum by turbidimetric immunoassay. Measurement of antigen-antibody reaction by the end-point method. Liquid reagents, ready for use. Measuring Range: 0 – 800 mg/dL. Detection Limit: 40 mg/dL

Auxiliary Reagents supplied separately

– saline solution 0.9g/dl

– set of calibrators to prepare the 5-level curve: IMT796-STD005 Protein standard set, or alternatively- high concentration calibrator, to generate a reference curve: IMT794-STD001 Protein standard high– CONTROLS: IMT805-CTR001 Himmunology control low

IMT806-CTR001 Immunology control high.

REF IMT287-K080  R1 2x25ml   R2 1x5ml

LIPASE

For the quantitative in vitro determination of Lipase in serum

and plasma. Turbidimetric method measured at 340 nm. Reagents, ready for use. Reaction is linear up to a concentration of  500 U/L.

REF IMT275-K080  R1a 1x70ml   R2b 20×2.5ml + CAL 4x3ml

MICROALBUMIN

Quantitative determination of  Microalbumin (MAL) in human urine by turbidimetric immunoassay.

Measurement of antigen-antibody reaction by the end-point method.

Liquid reagents, ready for use.

Measuring Range : 0 – 400 mg/dL  Detection Limit: 3 mg/dL.

REF IMT284-K045  R1 2x20ml   R2 1x5ml + CAL 1x1ml

REUMATOID FACTOR (RF)

Reagents for measurement of RF concentration.

Rheumatoid factors (RF) cause agglutination of the latex particles coated with human gammaglobulin. The agglutination of the latex particles is proportional to the RF concentration and can be measured by turbidimetry.

Linearity limit: 160 IU/mL. Detection limit: 6.9 IU/mL.

REF IMT482-K050  R1 2x20ml   R2 1x10ml + CAL 1x2ml

TRANSFERRIN

Quantitative determination of  Transferrin (TRF) in human serum by turbidimetric immunoassay.

Measurement of antigen-antibody reaction by the end-point method.

Liquid reagents, ready for use.

Measuring Range : 0 – 500 mg/dL  Detection Limit: 30 mg/dL.

REF IMT285-K080  R1 3x25ml   R2 1x5ml

IMMUNOTURBIDIMETRY    

STANDARDS & CALIBRATORS & CONTROLS

FIBRINOGEN STANDARD

Preparation of reference curves for quantitative immunochemical determination of Fibrinogen in citrated plasma by turbidimetry and nephelometry. A lyophilized preparation of stabilized, buffered human plasma containing less than 5% bovine plasma. After reconstitution, the Fibrinogen Standard is stable for 48 hours at +2 to +8 °C when stored in the original vial or in a sealed polypropylene container.

REF IMT780-STD000.5      1x5ml

FIBRINOGEN CONTROL

Accuracy control for the quantitative immunochemical determination of Fibrinogen in citrated plasma by turbidimetry and nephelometry.

A lyophilized preparation of stabilized, buffered human plasma containing less than 5% bovine plasma. After reconstitution, the Fibrinogen Control is stable for 48 hours at +2 to +8 °C when stored in the original vial or in a sealed polypropylene container.

REF IMT791-CTR001      1x1ml

PEDIATRIC STANDARD

Preparation of reference curves for quantitative immunochemical determination of proteins in human pediatric sera or urine. Pediatric standard is a defibrinated human plasma, liquid stabilised and filtered through 0.2 . After first opening the container, the serum can be used for 6 weeks if stored tightly closed at + 2 ° to 8 °C after use.

Do not freeze.

ASSIGNED VALUES: 2-Macroglobulin – Transferrin – IgA – IgG – IgM – Kappa Light Chain – Lambda Light Chain

REF IMT784-STD001      1x1ml

PROTEIN STANDART HIGH

Preparation of reference curves for quantitative immunochemical determination of proteins in human serum.

Protein standard serum is a defibrinated human plasma, liquid stabilised and filtered through 0.2 m.

ASSIGNED VALUES: Albumin, a1-Antitrypsin, a2-Macroglobulin, a1-Acidglycoprotein, Ceruloplasmin, Haptoglobin, Transferrin, IgA, IgG, IgM, Complement C3, Complement C4, Antithrombin III, C1 esterase inhibitor, Kappa Light Chain, Lambda Light Chain, Prealbumin.

REF IMT794-STD001      1x1ml

PROTEIN STANDARD SET

Preparation of reference curves for quantitative immunochemical determination of proteins in human serum.

Protein standard serum is a defibrinated human plasma, liquid stabilised and filtered through 0.2 m.

ASSIGNED VALUES: Albumin, a1-Antitrypsin, a2-Macroglobulin, a1-Acidglycoprotein, Ceruloplasmin, Haptoglobin, Transferrin, IgA, IgG, IgM, Complement C3, Complement C4, Antithrombin III, C1 esterase inhibitor, Kappa Light Chain, Lambda Light Chain, Prealbumin.

REF IMT796-STD005      5x1ml

HbA1c CALIBRATOR SET

Calibrator Set is a set of four liquid-stable calibrators with different levels based on human blood material (erythrocytes). After reconstitution the calibrator bottles should be stored at 2-8 ° C tightly closed. Stability is at least thirty days at 2-8 ° C.

REF IMT137-CAL002      1x2ml + 4×0.5ml

HbA1c CONTROL SET

Control set is a set of liquid-stabe controls with two levels: normal and elevated, suitable during the determination of glycated haemoglobin (HbA1c) in human blood by turbidimetry, nephelometry or enzymatic method. Stabilized and lyophilized hemolysates prepared from human erythrocytes.

REF IMT128-CTR001      2×0.5ml

IMMUNOLOGY CONTROL LOW

Accuracy control for the determination of proteins in serum by turbidimetry and nephelometry. After first opening the container, the serum can be used for 6 weeks if stored tightly closed at + 2 to + 8 °C after use. Do not freeze.

ASSIGNED VALUES: Albumin, 1-Acidglycoprotein, 1-Antitrypsin,

2-Macroglobulin, ASL, Antithrombin III, Β2-Microglobulin, Ceruloplasmin, Complement C3, Complement C4, C1 esterase inhibitor, CRP, Ferritin, Haptoglobin, IgA, IgG, IgM, Kappa Light Chain, Lambda Light Chain, Prealbumin, RF, Transferrin.

REF IMT805-CTR001      1x1ml

IMMUNOLOGY CONTROL HIGH

Accuracy control for the determination of proteins in serum by turbidimetry and nephelometry. After first opening the container, the serum can be used for 6 weeks if stored tightly closed at + 2 to + 8 °C after use. Do not freeze.

ASSIGNED VALUES: Albumin, 1-Acidglycoprotein, 1-Antitrypsin,

2-Macroglobulin, ASL, Antithrombin III, Β2-Microglobulin, Ceruloplasmin, Complement C3, Complement C4, C1 esterase inhibitor, CRP, Ferritin, Haptoglobin, IgA, IgG, IgM, Kappa Light Chain, Lambda Light Chain, Prealbumin, RF, Transferrin.

REF IMT806-CTR001      1x1ml

MICROALBUMIN CONTROL

Accuracy control for quantitative immunochemical determination of Microalbumin in urine by turbidimetry and nephelometry. A dilution of defibrinated human plasma with phosphate buffered saline, liquid stabilised and filtered through 0.2 m. After first opening the container, the control can be used for 6 weeks if stored tightly closed at + 2 ° to 8 °C after use. Do not freeze.

REF IMT790-CTR001      1x1ml

RHEUMATOID CONTROL SERUM LEV 1

The Rheumatoid Control Serum is a lyophilized human serum containing rheumatoid proteins in concentrations suitable for the quality control of the clinical laboratories. The components of the reconstituted Rheumatoid Control Serum are stable for at least 10 days at 2-8ºC and 6 months at –20ºC.

ASSIGNED VALUES – Latex-Turbidimetry: Anti-streptolisin O (ASO), Rheumatoid Factor (RF), C-Reactive Protein (CRP).

REF IMT798-CTR001      1x1ml

RHEUMATOID CONTROL SERUM LEV 2

The Rheumatoid Control Serum is a lyophilized human serum containing rheumatoid proteins in concentrations suitable for the quality control of the clinical laboratories. The components of the reconstituted Rheumatoid Control Serum are stable for at least 10 days at 2-8ºC and 6 months at –20ºC.

ASSIGNED VALUES – Latex-Turbidimetry: Anti-streptolisin O (ASO), Rheumatoid Factor (RF), C-Reactive Protein (CRP).

REF IMT799-CTR001      1x1ml